MAUDE Adverse Event Report: STRYKER-ENDOSCOPY PHOENIX USA; CATHETER, RECORDING, ELECTRODE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY PHOENIX USA; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 201101
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported catheter bending mechanism is not functioning properly and does not operate as intended.There was no patient injury or medical intervention and the extended procedure time reported was a few minutes.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.The complaint device was received in its secondary packaging with no restraints to properly secure it in place during shipping and handling.Upon visual inspection of the received complaint device, no relevant visible damage was detected.The catheter shaft, handle, steering mechanism, connector pins, and electrodes appeared to be intact.During functional inspection, the catheter deflected smoothly when the steering mechanism was actuated, indicating proper mechanical function.However, during deflection the distal tip appeared to be distorted when the catheter was fully deflected.Additionally, the device was evaluated and the device did not meet the planarity specification.The device inspection indicated that the "does not operate as intended¿ portion of the reported event was confirmed, as the curve of the device after deflection did not meet specification.The ¿bending mechanism is not functioning properly¿ portion was not confirmed, as the device demonstrated proper mechanical function of the handle.Since the device was return to stryker without the proper restraint to secure it in place during shipping, the extent of the issue experienced by the customer could not be fully determined, as shipping damage may have affected the condition of the returned device.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the likely root cause is a curve issue as a result of mishandling subsequent to distribution, including shipping and storage conditions or improper manipulation of the device.The instructions for use (ifu) state: - ep catheters should be used only by or under the supervision of an appropriately trained physician using proper procedures and techniques.- do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.- do not alter this device.- inspect the packaging and catheter for damage or defects prior to use.- avoid excessive torque, stretching, kinking and/or bending of catheter, as this may interfere with distal tip shaping or cause damage to internal electrode wires.- avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.Handle catheter with care to avoid improper electrical functioning.- do not insert or withdraw the catheter without straightening the catheter tip.- use fluoroscopic guidance during catheter positioning.- prior to use, store reprocessed ep catheters in a cool, dry, dark place.The reported event will continue to be monitored through post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER-ENDOSCOPY PHOENIX USA
10232 s. 51st st.
phoenix AZ 85044
Manufacturer (Section G)
STRYKER-ENDOSCOPY PHOENIX USA
10232 s. 51st st.
phoenix AZ 85044
Manufacturer Contact
arsalan rizvi
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key24670655
Report Number0002090040-2026-00006
Device Sequence Number5067325
Product Code NLH
UDI-Device Identifier00885825003845
UDI-Public00885825003845
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K030187
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number201101
Device Catalogue Number201101
Device Lot Number5579374
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2026
Initial Report FDA Received Date03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-