Model Number BF-H190
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was observed during the device inspection that the bronchovideoscope had an image with interference.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device was returned for inspection.Based on the results of the investigation and the information provided it is likely due to electrical/electronic component problem.The most probable cause is component failure.The root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24671435
Report Number9610595-2026-23480
Device Sequence Number18732036
Product Code EOQ
UDI-Device Identifier04953170434754
UDI-Public04953170434754
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K121959
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-H190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2026
Initial Report FDA Received Date03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-