| Model Number |
453944 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause could be determined.It should be noted that the ifu states that if the stent system was removed prior to expansion, do not re-insert as the stent and/ or the delivery system may have been damaged during the initial attempt to cross the lesion or during the withdrawal.
|
| |
|
Event or Problem Description
|
|
The orsiro mission drug-eluting stent system was selected for treatment of a mildly calcified lesion (90 % stenosis degree) in a segment of the distal lad.A pre-dilation of the lesion was performed.The affected device was introduced into the patient's body but could not cross the lesion.The target lesion was further pre-dilated but it was still not possible to cross.
|
| |
|