Model Number 310C29
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Material Split, Cut or Torn (4008); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Cusp Tear (2656); Mitral Valve Insufficiency/ Regurgitation (4451); Insufficient Information (4580)
Date of Event 02/27/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that 14 years post implant of this 29mm mitral bioprosthetic valve, it was explanted and replaced with a 27mm mitral bioprosthetic valve of the same model.The reason for the replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Updated data: a.4: weight in lbs: b.1: adv evnt/prod prob: b.2: outcome attributed to adverse event b.5: event description h.6: coding medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received that reported the reason for replacement as severe mitral regurgitation and a leaflet tear at the base of the commissure and annulus.It was further noted that a tricuspid valve repair was performed concomitantly.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC MITRAL BIOPROSTHETIC HEART VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24673361
Report Number2025587-2026-00325
Device Sequence Number13333385
Product Code LWR
UDI-Device Identifier00613994688927
UDI-Public(01)00613994688927(17)160120
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2016
Device Model Number310C29
Device Catalogue Number310C29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Supplement Date Received by Manufacturer03/24/2026
Initial Report FDA Received Date03/24/2026
Supplement Report FDA Received Date04/01/2026
Date Device Manufactured02/09/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
Patient Weight80 KG
-
-