• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number H1-M
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: product type: 0676-directional atherectomy; implant date n/a; explant date n/a.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during treatment of a 100% stenosis due to plaque in the right superficial femoral artery (proximal and mid segments, lesion length 30cm, artery diameter 6mm, minimal tortuosity and calcification), the atherectomy h1-m hawkone m experienced a mechanical jam and the cutter driver/thumbswitch stopped functioning while in use in the patient.It was possible to turn off the thumbswitch. the cutter was positioned outside the housing at the time of the issue and during device removal, with no deformation noted in the cutter.This issue is reported for 2 hawkone m atherectomy devices.A 6fr non-medtronic sheath and a non-medtronic guidewire were used.The vessel was pre-dilated with a 3x150 nanocross percutaneous transluminal angioplasty balloon.The vessel was post-dilated.No resistance was encountered when advancing the devices. the devices were safely removed from the patient.The procedure was completed with percutaneous transluminal angioplasty. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Image analysis the customer returned two video clips for evaluation 2567: this clip depicts the hawk cutter driver device and the thumb switch is being switched on and off, the audio of the file indicates that the device is intermittently being operational.2570: this clip depicts a fluoroscopic image of the hawk device in the patients vasculature, the audio of the file indicates that the device is intermittently being operational.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 6FR
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24673380
Report Number9612164-2026-01498
Device Sequence Number13333528
Product Code MCW
UDI-Device Identifier00643169968332
UDI-Public(01)00643169968332(17)281023(10)0013097247
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K161361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberH1-M
Device Lot Number0013097247
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2026
Supplement Date Received by Manufacturer04/01/2026
Initial Report FDA Received Date03/24/2026
Supplement Report FDA Received Date04/15/2026
Date Device Manufactured10/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PRODUCT ID H1-M (0013144806),
Patient SexMale
-
-