| Model Number |
UNIVERS VAULTLOCK GLENOID, SMALL |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Erosion (1750); Insufficient Information (4580)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On 04-mar-2026, a sales representative reported via email that a patient underwent a revision from eclipse + vaultlock tsa to rtsa on (b)(6) 2026.The patient¿s original surgery date was on (b)(6) 2024.The revision was performed due to subscapularis failure and not due to failure of the eclipse or vaultlock implants.However, significant vaultlock erosion was noted at less than two years post-implantation.The arthrex implants explanted during the revision included: ar-9106-01 arthrex univers vaultlock glenoid, small, ar-9301-02 arthrex eclipse cage screw medium (35.0 mm), ar-9301-39cpc arthrex eclipse trunion 39 mm, slotted, tps cap, and an ar-9339-18 arthrex eclipse humeral head, 39/18.Additional information was received on 10-mar-2026: the patient was initially treated on (b)(6) 2024 for a subscapularis tear sustained during a physical altercation.During the revision, anterior erosion measuring approximately 2¿3 mm was observed.The surgeon did not comment on whether the erosion contributed to the need for revision or if it was an incidental finding.The revision procedure was completed successfully with no case delay and no additional anesthesia reported.No further surgeries are anticipated.The patient was described as stable immediately following the revision procedure; postoperative status beyond leaving the room is unknown.Additional information was received on 16-mar-2026: the arthrex implant ar-9400-sbk eclipse speedscap implant system was explanted during the revision.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a patient-specific event.
|
| |
|
Search Alerts/Recalls
|
|
|