| Catalog Number |
05.001.240 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: section 10: general condition section 30: leakage test using bubble emission technique section 40: check coupling screw of saw blade coupling section 50: check saw blade coupling section 60: check saw head's locking mechanism section 80: check for sticky trigger section 140: check general function of device section 170: check oscillation frequency with frequency meter.During the service evaluation the following failures were identified: visual : component damaged - thread saw blade coupling internal finding : leak tightness test failure functional: will not run.Functional: sticky trigger.Conclusion: failure reported is confirmed.Root cause: improper maintenance.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the battery handpiece device had a sticky trigger.It was noted in the service order that the device had an unspecified malfunction and needed repair.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|