| Catalog Number |
IAB-S730C |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted, the balloon can not inflate completely, change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted, the balloon can not inflate completely, change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) the serial number on the returned sample is (b)(6).Returned for investigation was a 30cc 7fr rediguard intra-aortic balloon catheter (iabc) with the original packaging box that matches the serial number of the returned sample.The sample was returned in a cardboard box and was loosely packed within the original packaging carton.Returned with the sample was the supplied 30cc inflation driveline tubing; no damage or abnormalities were noted.Upon return, the peel away sheath was noted on the iabc outer lumen.The one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.Dried blood was noted on the exterior surfac-es of the returned sample.No obvious blood was noted within the helium pathway.No visual damage or abnormalities were noted to the returned sample.The bladder thickness was meas-ured at six points with measurements ranging from 0.0064in-0.0069in and was within specifica-tion of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.No abnormali-ties or debris were noted.The iabc was connected to an iabp with the returned 30cc inflation driveline tubing.The iabc was inserted into the "t" tube and 75mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The iabc was leak tested in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris was noted.A device history record (dhr) review was con-ducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint that the "balloon cannot inflate completely" is not confirmed.The returned iabc bladder was fully intact with no abnormalities noted.During the investigation, the iabc fully inflated, and no leaks/alarms were noted.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
|
| |
|
Search Alerts/Recalls
|