• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIGHT SCIENCES INC OMNI EDGE SURGICAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIGHT SCIENCES INC OMNI EDGE SURGICAL SYSTEM Back to Search Results
Catalog Number 1-112
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The physician who performed an omni surgery reported that the catheter detached from the cannula entirely.The catheter was removed from the eye using forceps and the procedure was completed.No patient injury had been reported and no other device was used.
 
Additional Manufacturer Narrative
The returned device was carefully evaluated and was determined that there is no evidence that the device design or manufacturing process was the cause of the incident.A review of device lot history and analysis of the most suspected components (i.E., cannula and catheter) concluded that there were no production non-conformances or any other anomalies that would have contributed to this event.Engineering had previously recreated the severing of catheter failure which did not point to a design or manufacturing issue as the cause of the failure.The mode of failure does suggest it's associated with a use error.The ifu provides cautionary statements related to possible kinking or damaging (e.G., severing) of microcatheter should the steps not be correctly performed.Therefore, it is concluded that the probable root cause for this event is likely due to a surgical technique or use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNI EDGE SURGICAL SYSTEM
Common Device Name
OMNI EDGE SURGICAL SYSTEM
Manufacturer (Section D)
SIGHT SCIENCES INC
4040 campbell ave
suite 100
menlo park CA 94025
Manufacturer (Section G)
SIGHT SCIENCES INC
4040 cmapbell ave
suite 100
menlo park CA 94025
Manufacturer Contact
ranjani madhavan
4040 campbell ave
suite 100
menlo park, CA 94025
MDR Report Key24677794
Report Number3010363671-2026-00001
Device Sequence Number10359952
Product Code MRH
UDI-Device Identifier00858027006594
UDI-Public(01)00858027006594(17)270811(10)F25H11-02
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K232214
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number1-112
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2026
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/24/2026
Supplement Report FDA Received Date04/23/2026
05/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-