Catalog Number D138505
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a cardiac ablation procedure with a vizigo and a blood leak from the hemostatic valve occurred.During the first insertion of the vizigo short, there was reverse blood from the hemostatic valve.The valve was not broken/cracked or totally separated.The hemostatic valve was not dislodged inside or outside of the hub and the brim cap/hub did not detach from the sheath.The issue occurred while the device was being used on the patient but no air entered the patient¿s body.Blood loss was a few drops.The issue was resolved by replacing the sheath.
 
Additional Manufacturer Narrative
On 8-apr-2026, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Additional Manufacturer Narrative
It was reported that a patient underwent a cardiac ablation procedure with a vizigo and a blood leak from the hemostatic valve occurred.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.The brim cap, hemostatic valve, and silicone ring were placed in their original place.Afterward, the sheath was blocked at the distal end and then pressurized at both positive and negative pressure, and no issues were observed.No dislodgement, air leaks, bubbles, or other failures with the hemostatic valve were observed during the test.A device history record evaluation was performed for the finished device lot number and no internal action was found during the review.The hemostatic valve leak reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: use direct imaging guidance (such as fluoroscopy or intracardiac ultrasound) to monitor the advancement of the catheter and removal of the catheter from the sheath.Move the catheter carefully to avoid cardiac damage, perforation, or tamponade.If resistance is encountered, do not use excessive force to advance or withdraw the catheter through the sheath.Use best practices for inserting or retracting any device at the hemostatic valve.Monitor any potential air entrapped and completely aspirate any observed air out of the side port.Do not remove dilator or catheter rapidly, damage to hemostatic valve may occur.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIZIGO
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
FREUDENBERG MEDICAL LLC
2301 centennial blvd
jeffersonville IN 47130
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key24678158
Report Number2029046-2026-00949
Device Sequence Number21350006
Product Code DYB
UDI-Device Identifier10846835026269
UDI-Public(01)10846835026269(17)271023(10)16804373
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K170997
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD138505
Device Lot Number16804373
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2026
Supplement Date Received by Manufacturer04/08/2026
04/12/2026
Initial Report FDA Received Date03/24/2026
Supplement Report FDA Received Date04/08/2026
04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-