| Catalog Number |
D138505 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cardiac ablation procedure with a vizigo and a blood leak from the hemostatic valve occurred.During the first insertion of the vizigo short, there was reverse blood from the hemostatic valve.The valve was not broken/cracked or totally separated.The hemostatic valve was not dislodged inside or outside of the hub and the brim cap/hub did not detach from the sheath.The issue occurred while the device was being used on the patient but no air entered the patient¿s body.Blood loss was a few drops.The issue was resolved by replacing the sheath.
|
| |
|
Additional Manufacturer Narrative
|
|
On 8-apr-2026, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent a cardiac ablation procedure with a vizigo and a blood leak from the hemostatic valve occurred.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.The brim cap, hemostatic valve, and silicone ring were placed in their original place.Afterward, the sheath was blocked at the distal end and then pressurized at both positive and negative pressure, and no issues were observed.No dislodgement, air leaks, bubbles, or other failures with the hemostatic valve were observed during the test.A device history record evaluation was performed for the finished device lot number and no internal action was found during the review.The hemostatic valve leak reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: use direct imaging guidance (such as fluoroscopy or intracardiac ultrasound) to monitor the advancement of the catheter and removal of the catheter from the sheath.Move the catheter carefully to avoid cardiac damage, perforation, or tamponade.If resistance is encountered, do not use excessive force to advance or withdraw the catheter through the sheath.Use best practices for inserting or retracting any device at the hemostatic valve.Monitor any potential air entrapped and completely aspirate any observed air out of the side port.Do not remove dilator or catheter rapidly, damage to hemostatic valve may occur.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|