| Model Number |
453934 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
The orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion (90 % stenosis degree) in a severely tortuous segment of the medial lad.A pre-dilation of the lesion was performed.The affected device was introduced into the patient's body but could not cross the lesion.
|
| |
|