MAUDE Adverse Event Report: SPECTRANETICS CORPORATION GLIDELIGHT LASER SHEATH; DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS CORPORATION GLIDELIGHT LASER SHEATH; DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS Back to Search Results
Model Number 500-302
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Unintended Radiation Exposure (4565); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
A lead extraction procedure commenced to remove a right ventricular (rv) and right atrial (ra) lead due to non-function.Spectranetics lld ez lead locking devices (lld ezs) were inserted into each lead to provide traction.Beginning with a spectranetics 14f glidelight laser sheath, the leads were heavily adhered, and during repeated advancement and retraction, laser light was observed exiting laterally from the catheter tip (mdr #3007284006-2026-00028).Therefore, a second glidelight laser sheath was utilized to remove the leads.The procedure was completed with no reported patient harm.The second glidelight was not reported by the facility, but was returned for evaluation on 17mar2026.Visual inspection found a breach to the outer jacket just distal to the bifurcate handle with broken fibers and melting observed (mdr #).Additionally, heavy biologics were present at the distal tip.This event is being reported for the second glidelight due to unintended radiation exposure, potential for harm.
 
Additional Manufacturer Narrative
B6/b7) patient information regarding relevant tests/laboratory data or medical history - not available from facility.H6) based on the device evaluation and investigation, this has been determined to be a use related failure.Historically, the manufacturer has seen this failure when the force required to advance and maneuver is greater than the force that the working length can accommodate.When these forces are applied to the side of the outer jacket, at or near the bifurcate handle, the device becomes kinked.Broken fibers in the area redirect the laser energy, which creates a breach in the outer jacket.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDELIGHT LASER SHEATH
Common Device Name
DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS
Manufacturer (Section D)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
nicole worford
9965 federal drive
colorado springs, CO 80921
719447-2465
MDR Report Key24680536
Report Number3007284006-2026-00091
Device Sequence Number15895306
Product Code MFA
UDI-Device Identifier00813132020330
UDI-Public(01)00813132020330(17)260215(10)FGB24B08A
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P960042
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/15/2026
Device Model Number500-302
Device Catalogue Number500-302
Device Lot NumberFGB24B08A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2026
Initial Report FDA Received Date03/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PHILIPS LASER SYSTEM.; SPECTRANETICS 14F GLIDELIGHT LASER SHEATH.; SPECTRANETICS LLD EZ LEAD LOCKING DEVICES.
Patient SexUnknown
-
-