| Catalog Number |
00630505036 |
| Medical Device Problem Codes |
Fracture (1260); Naturally Worn (2988); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately three years post-implantation, the patient underwent a right hip revision due to reported noises and premature poly wear.The patient reported squeaking during walking and mobilization without painful symptoms, which limited daily activities.The revision was performed and there was premature wear of the polyethylene.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - event occurred in italy.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately three years post-implantation, the patient underwent a right hip revision due to reported noises and premature poly wear and the poly was broken.The patient reported squeaking during walking and mobilization without painful symptoms, which limited daily activities.The revision was performed and there was premature wear of the polyethylene.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b5; d4; d9; h2; h3; h4; h6
complaint is confirmed.Visual examination of the returned product identified there is unusual wear on the i.D.Chamfer and top rim surface that remains on the device.Approximately half of top rim surface around the liner is fractured, 3 fractured fragments were returned with the device.The top rim remaining on the device is bent with a fractured section where the bend occurs.The post on the o.D.Has gouges and visually appears to be partially fractured.There are gouges and scratches present to the o.D., i.D., and top rim surfaces as well.No other damage was noted during visual evaluation.This device was implanted for approximately 3 years.Product identifier verification: no p.I.¿s are able to be verified due to damage.No dimensional evaluation performed due to damage.Fracture analysis performed, the timeline of the failure modes cannot be stated with certainty; however, it is possible that the liner became disassociated from the acetabular shell through a concurrent fracture of the boss and the rim, allowing the liner to bypass the locking mechanism.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Medical records were provided and reviewed by a health care professional/radiologist.The review identified right total hip arthroplasty initially showed the femoral head component well centered within the acetabular cup on.On follow-up imaging, the femoral head component is asymmetric within the acetabular cup superiorly, consistent with superior cup polyethylene wear or disruption.There is no gross evidence of loosening or osteolysis.No specific risk factor for this complication is identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|