| Model Number |
BF-1T180 |
| Medical Device Problem Codes |
Degraded (1153); Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed during the device evaluation, the bronchovideoscope exhibited the connecting tube had the coating peeling and the distal end had a burned.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: component failure, environmental temperature and degradation problems identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|