| Catalog Number |
SSX-650 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture, baker grade iii".This record is for the right side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b3, b5, d6b, h6.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture, baker grade iii".This record is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis results:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (deformation and crease) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, corrected and/or changed data: d.4, d.9, h.3, h.6.
|
| |
|