| Model Number |
IRES WARMER |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.Legal manufacturer: gehc trout - 3114 n grandview blvd.Usa waukesha, wi 53188.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a malfunction resulting in loss of oxygen therapy.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The o2 regulator was replaced to resolve the issue.
|
| |
|