| Model Number |
CI-1600-05 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was sliced and tool damage to the silicone overmold on the top and bottom cover of the device.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed broken electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The device passed some of the electrical tests performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
The recipient reportedly experienced decreased performance.A review of the recipient's test data indicates impedance issues.Programming adjustments have been made; however, the issue did not resolve.The recipient's device was explanted.The recipient was re-implanted with another cochlear device.
|
| |
|