WILSON-COOK MEDICAL INC (COOK ENDOSCOPY) HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC; DILATOR, ESOPHAGEAL
|
Back to Search Results |
|
| Model Number |
G51799 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an upper endoscopic dilation procedure, the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.It was reported that there was a leak in the distal end of the balloon.The procedure was completed with another cook [hbd-w-10-11-12] dilation balloon.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report due to a pinhole in the balloon material.The device was returned with the balloon deflated and an unknown clear substance within the balloon.A cook dilation syringe (ds-60cc-s) was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was attempted to be inflated.The balloon would fully inflate; however, it would not hold pressure, and leakage was observed from a pinhole in the proximal area of the balloon material.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis note: the pre-loaded wire guide and stem bumper associated with this device were not included in the return.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the returned device confirmed the report due to a pinhole in the balloon material.A definitive cause for the reported observation could not be determined.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.In the report it does not state if negative pressure and lubrication were applied to the balloon prior to advancing down the endoscope accessory channel.A possible contributing factor to a leakage is failure to apply negative pressure and lubrication to the balloon prior to advancement.The instructions for use advise the user: "to facilitate passage through the endoscope, apply negative pressure to the catheter." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use also advise the user: "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." the instructions for use also advises the user: "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel." this activity will also aid in endoscopic advancement and balloon preservation.A leakage/damage to the balloon material can occur if the balloon comes into contact with a sharp object or a burr in the endoscope channel.Another possible contributing factor to a leakage in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage wire guided balloons esophageal-pyloric-colonic are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment, no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report due to a pinhole in the balloon material.The device was returned with the balloon deflated and an unknown clear substance within the balloon.A cook dilation syringe (ds-60cc-s) was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was attempted to be inflated.The balloon would fully inflate; however, it would not hold pressure, and leakage was observed from a pinhole in the proximal area of the balloon material.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.Note: the pre-loaded wire guide and stem bumper associated with this device were not included in the return.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the returned device confirmed the report due to a pinhole in the balloon material.A definitive cause for the reported observation could not be determined.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.In the report it states that it is unknown if negative pressure was applied to the balloon prior to advancing down the endoscope accessory channel.A possible contributing factor to a leakage is failure to apply negative pressure to the balloon prior to advancement.The instructions for use advise the user: "to facilitate passage through the endoscope, apply negative pressure to the catheter." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use also advise the user: "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." this activity will also aid in endoscopic advancement and balloon preservation.A leakage/damage to the balloon material can occur if the balloon comes into contact with a sharp object or a burr in the endoscope channel.Another possible contributing factor to a leakage in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage wire guided balloons esophageal-pyloric-colonic are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment, no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Event or Problem Description
|
|
During an upper endoscopic dilation procedure in the esophagus, the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.It was reported that there was a leak in the distal end of the balloon.The procedure was completed with another cook [hbd-w-10-11-12] dilation balloon.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|