• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ONE PIECE MAIN PUMP&PATIENT EXT TBG; ARTHROSCOPE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ONE PIECE MAIN PUMP&PATIENT EXT TBG; ARTHROSCOPE AND ACCESSORIES Back to Search Results
Model Number ONE PIECE MAIN PUMP&PATIENT EXT TBG
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during a posterior and anterior cruciate ligament surgery the inflow tubes of the arthroscopic pump became overpressurized and broke.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.This line was created for the tubing that was used with the reported device.Update 09-mar-2026: the tubing that was used with the pump are the ar-6415 lot: n5032.The pump pressure was set at 60 mmhg pressure.No alarm or error message received during the use just an unusual high pressure and continuous flow.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONE PIECE MAIN PUMP&PATIENT EXT TBG
Common Device Name
ARTHROSCOPE AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key24688260
Report Number1220246-2026-01635
Device Sequence Number15866476
Product Code HRX
UDI-Device Identifier00888867039131
UDI-Public(01)00888867039131(11)250617(17)300531(10)N5032
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K083707
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberONE PIECE MAIN PUMP&PATIENT EXT TBG
Device Catalogue NumberAR-6415
Device Lot NumberN5032
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2026
Initial Report FDA Received Date03/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-