| Model Number |
EVOLUTPRO-29-US |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Codes |
Thrombosis/Thrombus (4440); Insufficient Information (4580)
|
| Date of Event |
03/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that 6 years and 11 months following the implant of a transcatheter aortic valve, an unspecified valve failure occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Image review: the executive summary and a video sweep through the evolut were provided.There appears to be plaque/ thrombus inside the evolut frame.The evolut leaflets also appear mildly thickened.No calcification was noted on the leaflets.Updated data: b1.B2.B5.H1.H6.Additional codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the failure was due to thrombus formation as it was noted that there was a tear between the left and non-commissure leaflets.
|
| |
|
Search Alerts/Recalls
|