MEDTRONIC MEXICO S. DE R.L. DE CV EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVOLUTPRO-29-US
Medical Device Problem Codes Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001); Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Thrombosis/Thrombus (4440); Insufficient Information (4580)
Date of Event 03/19/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that 6 years and 11 months following the implant of a transcatheter aortic valve, an unspecified valve failure occurred.
 
Additional Manufacturer Narrative
Image review: the executive summary and a video sweep through the evolut were provided.There appears to be plaque/ thrombus inside the evolut frame.The evolut leaflets also appear mildly thickened.No calcification was noted on the leaflets.Updated data: b1.B2.B5.H1.H6.Additional codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received that the failure was due to thrombus formation as it was noted that there was a tear between the left and non-commissure leaflets.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO TRANSCATHETER AORTIC VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24688720
Report Number9617601-2026-02016
Device Sequence Number15866365
Product Code NPT
UDI-Device Identifier00763000017835
UDI-Public(01)00763000017835(17)210129
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2021
Device Model NumberEVOLUTPRO-29-US
Device Catalogue NumberEVOLUTPRO-29-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2026
Supplement Date Received by Manufacturer04/03/2026
Initial Report FDA Received Date03/25/2026
Supplement Report FDA Received Date04/28/2026
Date Device Manufactured02/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age86 YR
Patient SexMale
Patient Weight62 KG
-
-