| Model Number |
WAIR130A |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to olympus for inspection, however the alleged failure was confirmed vis information from field service engineer (fse).A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause of chipped distal lens is traced to component failure due to degradation.The date of event is unknown.A supplement report will be submitted if provided.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid endoscope telescope exhibited chipped distal lens.There was no patient involvement.
|
| |
|