| Model Number |
BF-1T60 |
| Medical Device Problem Codes |
Corroded (1131); Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.In addition, the investigation found that the forceps channel port has corrosion.Based on the results of the investigation, the most probable cause suggests that due to some kind of stress such as damage or deterioration of parts.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the oes bronchofiberscope forceps channel port has corrosion.There was no patient involvement.
|
| |
|