| Catalog Number |
072227 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that a few days ago, at the facility, staff was attempted to remove a wound drain from a patient.The drain snapped off into the patient.They had to take the patient to the operating room for removal of the retained drain piece.They unsure of what they did with the drain piece whether that was sent to pathology for evaluation or just discarded.They just wanted to report that incident in case other have reported a similar issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: warnings: an effective closed suction drain system requires maintenance of the system to preserve patency.The drain must not be allowed to occlude nor the reservoir to completely fill; and reservoir suction must be maintained in order for the system to function properly.Verify that the system is functioning properly.If the system is not maintained properly, surgical complications, including hematomas, may result.Drain perforations or channels must lie within the wound or cavity to be drained, otherwise inadequate drainage may result.To avoid the possibility of drain damage or breakage: avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patient's degree of intolerance to any foreign object in the body.Care must be exercised to avoid damage to the drain.The tubing should be repeatedly checked during closure for free motion to avoid breakage and or fragment retention within the wound.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Corrections: d, e.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that a few days ago, at the facility, staff was attempted to remove a wound drain from a patient.The drain snapped off into the patient.They had to take the patient to the operating room for removal of the retained drain piece.They unsure of what they did with the drain piece whether that was sent to pathology for evaluation or just discarded.They just wanted to report that incident in case other have reported a similar issue.
|
| |
|
Event or Problem Description
|
|
It was reported that a few days ago, at the facility, staff was attempted to remove a wound drain from a patient.The drain snapped off into the patient.They had to take the patient to the operating room for removal of the retained drain piece.They unsure of what they did with the drain piece whether that was sent to pathology for evaluation or just discarded.They just wanted to report that incident in case other have reported a similar issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further review, bd has determined that this mdr was reported in error as it was found to be a duplicate of an event previously reported.The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|