DEPUY MITEK LLC US FMS; ARTHROSCOPE Back to Search Results
Catalog Number 284508
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopic rotator cuff repair procedure that the inflow tubing fms vue device 5 minutes after the start of the surgery, a water leak was discovered from the fms inflow section.Upon removing the inflow tube cover, a crack was found in the tube.The irrigation water was stopped, the inflow tube was removed and replaced with a new one.Another device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.All available information has been disclosed.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated, and no conclusion can be drawn at this time.E1: the reporter¿s complete facility address was not provided.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: it was reported that during an arthroscopic rotator cuff repair procedure, approximately 5 minutes after the start of the surgery, a water leak was discovered from the fms inflow section.Upon removing the inflow tube cover, a crack was found in the inflow tubing fms device.The irrigation water was stopped, the inflow tube was removed and replaced with a new one.The surrounding wet areas were wiped with gauze to remove moisture, the tube was set up according to the procedure, and the run button was pressed again.Water began to flow, but the low pressure indicator flashed on the fms vue ii pump-shaver box device without detecting pressure.The tube was reattached, and the joints of each tube and the fms connection points were checked.The power was turned off the fms, the tubing was replaced again with a new one, set up according to the procedure, and reattached.It then operated normally.The initial fms setup was performed by the circulating nurse before entering the room, and it was confirmed that there was no damage to the tube.There was no surgical delay and no harm to the patient.No further information is available.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the complaint device was not returned to depuy synthes, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (3011614), and there was a non-conformance unrelated to the reported complaint.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing record evaluation was performed for the finished device lot number, and there was a non-conformance unrelated to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key24690688
Report Number1221934-2026-01231
Device Sequence Number21259203
Product Code HRX
UDI-Device Identifier10886705023066
UDI-Public(01)10886705023066(10)3011614(17)260530
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951843
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2026
Device Catalogue Number284508
Device Lot Number3011614
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2026
Supplement Date Received by Manufacturer03/19/2026
04/29/2026
Initial Report FDA Received Date03/25/2026
Supplement Report FDA Received Date04/16/2026
05/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-