SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72201491 |
| Medical Device Problem Code |
Firing Problem (4011)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a meniscus repair, both ts of the ultra-fast-fix fell off at the same time as the first deployment occurred.The ts were not deployed through the meniscus.The procedure was resumed, after a non-significant delay, using an equivalent s+n back-up.The patient status is fine.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: case-(b)(4).H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The t1, t2, and suture were not returned.The variable depth straw was not returned.Bio debris is present.A functional evaluation could not be performed because both implants have already been deployed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|