| Model Number |
2N8371 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Rupture (1546); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified quantity of non-dehp y-type catheter extension sets came apart or ruptured during patient use with computed tomography (ct) contrast.The sets came apart or leaked either after tightening the caps (clearlink) or at the y-site (junction).There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h4, h6 (update codes), h11.H11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were not received for evaluation and no photos were provided showing the sets; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|