| Catalog Number |
MY8060 |
| Medical Device Problem Codes |
Leak/Splash (1354); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E.1.Address was not located, and il was used.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd texium needle-free syringe had connection issues.The following information was provided by the initial reporter: leakage occurs during connection and/or while pushing drug into the bag.Staff report the connection may not thread correctly with the bag access device.Notably, when the texium syringe is secured/tightened fully, leakage appears more likely, and slightly loosening may reduce leakage.Leakage has occurred with both 20 ml and 50 ml sizes and across different lots.Describe patient / hcp / user impact: hazardous drug exposure, chemo leak.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: due to no sample being received, an investigation could not be performed, and a root cause could not be determined.No product will be returned per customer.No investigation was performed.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Search Alerts/Recalls
|