• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION ULTRAVIT PROBE PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION ULTRAVIT PROBE PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752134
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that one of the three tubes dropped from the unit immediately after use during surgery.The procedure type was unknown.The procedure was completed after replacement of product with new one.There was no patient impact.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The used vitrectomy cutter was visually inspected.The black tube was found detached from the engine manifold of the cutter.Microscopic inspection revealed that a small amount of solvent had been applied around the tubing, preventing it from adhering to the wall of the manifold.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.The investigation determined that the most probable root cause of the customer¿s complaint was an assembly error that occurred during manufacturing of the probe manifold for this specific product configuration.During assembly, the tubing that forms part of the probe manifold is required to be wetted with a specified amount of solvent to ensure proper adhesion and mechanical bonding.Evidence collected during the returned product evaluation¿specifically the detachment of the gray line from the probe manifold¿indicates that insufficient solvent was applied during the wetting step.When an inadequate amount of solvent is used, the tubing surface is not fully activated, resulting in reduced bonding strength.Over time, or when subjected to normal operational stresses, this insufficient bond can lead to detachment of the manifold components.The failure mode observed on the returned product is consistent with incomplete wetting or improper application of solvent during assembly.Since no deviations, processing anomalies, or material defects were identified during the device record review for the reported lot, the issue is considered isolated to the manual wetting process step for this specific probe manifold product type.This process involves operator dependent tasks, and the variation in solvent application is likely attributable to a single unit assembly error rather than a systemic manufacturing issue.No evidence suggests that other units within the lot or other product lines are affected.This complaint has been reviewed, and it is determined that no further actions will be pursued at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION ULTRAVIT PROBE PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24696602
Report Number1644019-2026-01713
Device Sequence Number15979880
Product Code LRO
UDI-Device Identifier00380657521340
UDI-Public(01)00380657521340(17)270430(11)250505(10)17DPPU
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752134
Device Lot Number17DPPU
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/03/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/02/2026
Supplement Date Received by Manufacturer03/30/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/05/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUSTOM-PAK SURGICAL PROCEDURE PACK.
Patient SexUnknown
-
-