• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LEGION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LEGION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753075
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that during surgery, the ophthalmic system was not building vacuum in irrigation and aspiration mode, as well as in phacoemulsification mode.The surgery was completed on the same day by using another machine.There was no patient impact.
 
Additional Manufacturer Narrative
The company representative was able to replicate the reported issue.The nonconforming fluidics module was replaced to address this issue.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.There were no similar complaints reported for the ¿product serial¿ under investigation, with the same event code.The root cause of the reported event is attributed to the nonconforming fluidics module.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: 2026-17476.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION SYSTEM
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24697402
Report Number2028159-2026-00384
Device Sequence Number10109931
Product Code HQC
UDI-Device Identifier00380657530755
UDI-Public(01)00380657530755(17)350607(11)250609(10)17EVYA
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K191650
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065753075
Device Lot Number17EVYA
Was Device Available for Evaluation? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/03/2026
Supplement Date Received by Manufacturer05/19/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-