| Catalog Number |
LXMC17 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Dysphagia/ Odynophagia (1815)
|
| Date of Event |
03/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/26/2026.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? on what date did the implant take place? on what date did the explant take place? what is the lot number of the linx device? when using the linx sizing device what technique was used to determine the size? did the patient have an autoimmune disease? is the patient currently taking steroids / immunosuppressive drugs? did the patient have any pre-existing dysphagia or other conditions (other than gerd)? how severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? was there any hiatal or crural repair done at the same time as the implant? were there any other contributing factors that led to the removal of the device other than the reported dysphagia? besides the reported dysphagia, what was the reason for removal of the linx device? was the device found in the correct position/geometry at the time of removal? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that linx was explanted.Patient was having very bad dysphasia, dilated multiple times and still was having problems.Needed to explant.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 4/22/2026.Additional information was requested, and the following was obtained: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? on what date did the implant take place? on what date did the explant take place? what is the lot number of the linx device? when using the linx sizing device what technique was used to determine the size? did the patient have an autoimmune disease? is the patient currently taking steroids / immunosuppressive drugs? did the patient have any pre-existing dysphagia or other conditions (other than gerd)? how severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? was there any hiatal or crural repair done at the same time as the implant? were there any other contributing factors that led to the removal of the device other than the reported dysphagia? besides the reported dysphagia, what was the reason for removal of the linx device? was the device found in the correct position/geometry at the time of removal? i do not have any additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/28/2026.Investigation summary: a 17 beads linx device was returned in used condition.The device was returned in a locked position, in addition an attempt to unlocked was successfully made.A cut wire, bent wires and tooling marks were noted in some beads.Overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, tooling marks were noted in some beads.Link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.The device lot number is unknown; therefore, the device history record could not be reviewed.
|
| |
|
Search Alerts/Recalls
|