| Model Number |
H74939135152010 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the superficial femoral artery.A 1.5mm x 20mm x 143cm coyote es balloon catheter was advanced for dilation.During multiple inflations at nominal pressure for several seconds, the balloon ruptured.The device was removed using normal method without issue and the procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: device technical analysis upon receipt at our post market quality assurance laboratory, the outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 14mm from the tip.No other issues were identified during the product analysis.Device history record (dhr) review it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations that could have contributed to the reported event.Labeling review review of the instructions for use (ifu) confirmed that the content was sufficient and did not contribute to the reported event; therefore, no updates are required to the document at this time.Investigation conclusion based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the superficial femoral artery.A 1.5mm x 20mm x 143cm coyote es balloon catheter was advanced for dilation.During multiple inflations at nominal pressure for several seconds, the balloon ruptured.The device was removed using normal method without issue and the procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|
Search Alerts/Recalls
|