| Model Number |
72404014 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Nonstandard Device (1420); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404310.Serial number: n/a.Batch/lot number: 1100776414.Model/catalog description: pump ms.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient visited the hospital because his inflatable penile prosthesis (ipp) was not functioning properly.The patient underwent surgery two weeks later and upon inspection, a hole was identified in the left cylinder.The left cylinder and the pump were explanted and replaced.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient visited the hospital because his inflatable penile prosthesis (ipp) was not functioning properly.The patient underwent surgery two weeks later and upon inspection, a hole was identified in the left cylinder.The left cylinder and the pump were explanted and replaced.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to b1 adverse event/product problem.Model number/catalog number: 72404310
serial number: n/a
batch/lot number: 1100776414
model/catalog description: pump ms
unique identifier (udi) #: (b)(4).Upon receipt at our quality assurance laboratory, this inflatable penile prosthesis (ipp) underwent a thorough analysis.The cylinders were microscopically inspected and leak tested.Microscope inspection revealed that a cylinder had a hole in the middle from a sharp instrument.Leakage testing revealed a leak in a cylinder from a sharp instrument in the middle of the cylinder.The pump was microscopically inspected and functionally and leak tested.Microscopic inspection revealed that the pump tubing was cut down to the pump block; the tubing was not returned for analysis.The pump failed the activation tests for the functional test.The pump passed the leak testing.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of mechanical malfunction (leaks, incomplete inflation/ deflation, kinking) was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available and analysis results, the failure to pass the activation test identified in the pump could have caused or contributed to the reported clinical observation of mechanical malfunction.
|
| |
|