Model Number 72404014
Medical Device Problem Codes Mechanical Problem (1384); Nonstandard Device (1420); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Model number/catalog number: 72404310.Serial number: n/a.Batch/lot number: 1100776414.Model/catalog description: pump ms.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.
 
Event or Problem Description
It was reported that the patient visited the hospital because his inflatable penile prosthesis (ipp) was not functioning properly.The patient underwent surgery two weeks later and upon inspection, a hole was identified in the left cylinder.The left cylinder and the pump were explanted and replaced.There were no patient complications reported.
 
Event or Problem Description
It was reported that the patient visited the hospital because his inflatable penile prosthesis (ipp) was not functioning properly.The patient underwent surgery two weeks later and upon inspection, a hole was identified in the left cylinder.The left cylinder and the pump were explanted and replaced.There were no patient complications reported.
 
Additional Manufacturer Narrative
Correction to b1 adverse event/product problem.Model number/catalog number: 72404310 serial number: n/a batch/lot number: 1100776414 model/catalog description: pump ms unique identifier (udi) #: (b)(4).Upon receipt at our quality assurance laboratory, this inflatable penile prosthesis (ipp) underwent a thorough analysis.The cylinders were microscopically inspected and leak tested.Microscope inspection revealed that a cylinder had a hole in the middle from a sharp instrument.Leakage testing revealed a leak in a cylinder from a sharp instrument in the middle of the cylinder.The pump was microscopically inspected and functionally and leak tested.Microscopic inspection revealed that the pump tubing was cut down to the pump block; the tubing was not returned for analysis.The pump failed the activation tests for the functional test.The pump passed the leak testing.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of mechanical malfunction (leaks, incomplete inflation/ deflation, kinking) was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available and analysis results, the failure to pass the activation test identified in the pump could have caused or contributed to the reported clinical observation of mechanical malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CXR
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key24698471
Report Number2124215-2026-16558
Device Sequence Number10440125
Product Code JCW
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number72404014
Device Catalogue Number72404014
Device Lot Number1100584737
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2026
Supplement Date Received by Manufacturer05/21/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date06/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexMale
-
-