ALLERGAN (COSTA RICA) Style 410 Cohesive Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number FF-410375
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "intracapsular rupture".This record is for the left side.Device was explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported "intracapsular rupture".This record is for the left side.Device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b.5., d.6b., h.6.
 
Additional Manufacturer Narrative
Device evaluation: the device related to the reported events rupture received on june 15,2026, with lot number 2403267.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: - rupture: not observed through visual inspection.None of the other observations performed during the device analysis (deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
 
Event or Problem Description
Healthcare professional reported "intracapsular rupture".This record is for the left side.Device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Style 410 Cohesive Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key24698864
Report Number9617229-2026-05397
Device Sequence Number13289352
Product Code FTR
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/18/2017
Device Catalogue NumberFF-410375
Device Lot Number2403267
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/15/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2026
Supplement Date Received by Manufacturer04/28/2026
07/01/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date05/26/2026
07/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-