DEPUY MITEK LLC US FMS; ARTHROSCOPE Back to Search Results
Catalog Number 283459
Medical Device Problem Codes Mechanical Jam (2983); Output Problem (3005); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that it comes in used condition.The inner shaft was found seized.It was not possible to spin to any direction.No anomalies were found on the black connector.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the 4.0mm round burr plus 5pk would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to higher speeds applied on the handpiece which can increase the temperature of the device's shaft, lack of adequate irrigation and excessive side loading can contribute to a seized inner shaft, as per the instructions for use: adequate irrigation to the tip of the blade or burr is required to cool the blade and prevent accumulation of excised material in the surgical site.Excessive side loading may result in blade wear and degradation, as well as clogging and/or seizing.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing record evaluation was performed for the finished device lot number (m2409003), and no non-conformance was identified.
 
Event or Problem Description
It was reported that during device evaluation, it was determined that the inner shaft of the 4.0mm round burr device was seized, it was not possible to spin to any direction.It was noted in the service order that during a arthroscopic rotator cuff repair (arcr) for rotator cuff tears (rct) of the shoulder, the shaver had made a strange noise while in use, had stopped working, and was no longer usable.Even after changing the handpiece, it still wouldn't work, and even after changing the shaver tip, it stopped working again.The third shaver was still usable until the latter half of the operation.There were no surgical delay and no harm to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key24699154
Report Number1221934-2026-01253
Device Sequence Number10557156
Product Code HRX
UDI-Device Identifier10886705025428
UDI-Public(01)10886705025428(10)M2409003(17)290831
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number283459
Device Lot NumberM2409003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2026
Initial Report FDA Received Date03/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-