A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A trilogy 200 ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at the third-party service center, a service required alarm condition indicating the internal battery was in a low state of health was observed in the device's downloaded event log.The internal battery needs to be replaced to address the issue.The device was requested to return unrepaired to the customer.The sound abatement foam was replaced preventatively.The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|