• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA SCREW-IN LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA SCREW-IN LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5071-53
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Codes Stroke/CVA (1770); Loss of consciousness (2418)
Date of Event 02/09/2026
Type of Reportable Event Death
Event or Problem Description
It was reported that the right ventricular (rv) epicardial lead exhibited loss of capture on the current electrograms (egm), oversensing with sensed events noted post ventricular blanking, and oversensing of noise resulting in inhibition of pacing.It was noted that the patient had an altered mental status and was unresponsive.The rv epicardial lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 5071-53 lead, implanted (b)(6) 2019. dtpa2d1 crt-d, implanted (b)(6) 2023. 6725 adaptor, implanted (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Correction: b5 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right ventricular (rv) epicardial leads exhibited loss of capture on the current electrograms (egm), oversensing with sensed events noted post ventricular blanking, and oversensing of noise resulting in inhibition of pacing.It was noted that the patient had an altered mental status and was unresponsive.It was further reported that approximately a month prior, the cardiac resynchronization therapy defibrillator (crt-d) experienced electromagnetic interference from an external, unknown source.The crt-d remains in use.The crt-d and rv epicardial leads remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the patient was paced below the lower programmed rate due to the right ventricular (rv) leads sensing outside of the blanking period.Additionally, the patient experienced a stroke and ultimately passed away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW-IN LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24700200
Report Number2649622-2026-08484
Device Sequence Number10467982
Product Code DTB
UDI-Device Identifier00643169638600
UDI-Public(01)00643169638600(17)230122
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P120017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/22/2023
Device Model Number5071-53
Device Catalogue Number5071-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2026
Supplement Date Received by Manufacturer04/06/2026
05/05/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date05/05/2026
05/29/2026
Date Device Manufactured02/04/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5071-53 LEAD, 5071-53 LEAD, 6721M-50 LEAD
Outcome Attributed to Adverse Event Life Threatening; Death;
Patient Age74 YR
Patient SexFemale
Patient Weight61 KG
-
-