Per the field clinical specialist, during a transcatheter aortic valve replacement (tavr) procedure using a 23mm sapien 3 ultra resilia valve, during inflation there was too much forward pressure on commander system and advanced the valve ventricular resulting in a final position of 50/50.The perceived root cause of the valve movement was reported to be excessive forward pressure applied to the commander delivery system.Due to the suboptimal position of the first valve, a second valve was implanted successfully, while the first valve remained in situ.During deployment of the second valve, a balloon rupture of the second commander delivery system was observed and post dilated with a 22mm non-edwards balloon.Despite this additional event, the second valve was successfully implanted.The patient remained in stable condition following the procedure.
|
|
Per the instructions for use (ifu), valve malposition requiring intervention is a known potential adverse event associated with the transcatheter valve replacement (tvr) procedure.There are multiple patient and procedural factors that alone or in combination can cause or contribute to malposition, including improper positioning before deployment, poor image intensifier angle, poor coaxial alignment of the valve and delivery system, severe septal hypertrophy, minimally or bulky/severely calcified leaflets, preserved ejection fraction, significant landing zone calcification, loss of pacing capture, rapid deployment, the release of stored tension during deployment, and movement of the delivery system by the operator.The ifu cautions that incorrect sizing of the valve may lead to paravalvular leak, migration, or embolization.The device training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Physicians are extensively trained by edwards before they are qualified to use the sapien transcatheter heart valves (all models).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.The correct alignment and positioning of the device at the point of deployment is emphasized as a key factor in the placement and fixation of the device.Operators are also instructed to use fluoroscopy as the primary method of visualization for positioning and deployment.In patients with high-risk anatomical features for malposition (i.E., minimal leaflet calcification, severe septal hypertrophy), balloon valvuloplasty may indicate potential balloon movement during valve deployment.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate that procedural factors (movement of the delivery system caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|