| Model Number |
BI70002000 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(h3,h6): manufacturing representative went to the site to test the system, navigation inaccuracy identified.Recalibration with successful navigation parameter.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: m021083c001, serial/lot #: 4.2.1, ubd: , udi#: medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that failed a work order to indicate that the system was recalibrated due to an inaccuracy.There was no patient present.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000218 , product id: bi71000219, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information updated in b5.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received, representative recalibrated the image acquisition system(ias) after checking verification scans and found that they were not accurate.The ias active spheres needed to be recalibrated.Ias had been recalibrated.They have to use a navigation system to recalibrate the active trackers.There was no patient present.
|
| |
|