• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL DEVICES S.A. DE C.V. INSET II; UNO INSET II 60/6 SC1 BLUE MIMX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL DEVICES S.A. DE C.V. INSET II; UNO INSET II 60/6 SC1 BLUE MIMX Back to Search Results
Model Number MMT-943A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: patient city: (b)(6).Patient country: the united states.Unomedical hereby submits this initial report as part of its complaint remediation activities conducted under a corrective and preventive action (capa)/fda action plan.This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this initial information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged based on the review completed on 25-feb-2026 and investigation completed on 18-mar-2026 this medical device report (mdr) is being submitted as the initial report.Complaint investigation results: electronic quality management system (eqms) search: a query was run in the eqms on 18/mar/2026 against lot number 5362155 and similar malfunction codes: leakage from infusion site (cannula base part/cannula) (specific cause not identified), insertion site egress - trace moisture / superficial wetness.The review confirmed that lot 5362155 and the identified failure mode are not associated with any non conformance report (ncrs) or capas of the same or similar nature.Similar complaints search: a query was run in the eqms on 18/mar/2026 against lot number criteria equal 5362155 and similar malfunction codes: leakage from infusion site (cannula base part/cannula) (specific cause not identified), insertion site egress - trace moisture / superficial wetness.The number of complaints is (b)(4).The complaint numbers are complaint (b)(4) and (b)(4).Device history record (dhr) review: the lot 5362155 was manufactured according to work instruction (wi) version 28 and manufactured in the line 6, on 04/aug/2021 with a total of (b)(4) units.The device history record (dhr) was reviewed in accordance with applicable procedures.All required in-process and final tests were completed and met specified requirements.No deviations were identified, and no maintenance events were recorded that relate to the complaint code.Conclusion: dhr review supports compliance with manufacturing and quality requirements no issues noted.Visual evidence review: no photo was provided to support visual confirmation of the reported issue.Conclusion: unable to perform visual verification assessment based on available documentation only.Capa determination assessment - conclusion summary of complaint investigation: the investigation for this complaint was previously completed on 29-sep 2023 under child complaint investigation record (b)(4).The investigation has been revised to reflect the updated malfunction code and new reportability requirements.The original investigation remains valid and has not been modified.As part of this reassessment, an updated eqms search was performed which identified two related complaints for lot 5362155.No ncrs, capas, trends, or systemic issues were found.As required under the updated reportability criteria, a full device history record (dhr) review was conducted.All in process and final inspections met specification requirements, and no deviations, rework activities, or maintenance events associated with the malfunction were identified.No visual evidence was available to support confirmation of the reported issue.No photo evidence was provided, so the assessment was based on documentation only.Based on the expanded investigation including updated eqms searches and the dhr review no manufacturing or quality issues were identified.The record will be closed and monitored through tracking and trending per wi (monthly trips and alerts) root cause of problem: n/a - this complaint did not require escalation to capa therefore, no further root cause investigation has been completed.Corrective action as a result of the investigation: n/a - this complaint did not require escalation to capa therefore, no capa plan is available.Capa determination = no capa required.Complaint unconfirmed: following investigation, the malfunction remains unconfirmed for this complaint.
 
Event or Problem Description
Reference number (b)(4).Event occurred in the united states.It was reported that the patient experienced leakage at the site on 18-sep-2023, after the infusion set had been in use for 5 minutes.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET II
Common Device Name
UNO INSET II 60/6 SC1 BLUE MIMX
Manufacturer (Section D)
UNOMEDICAL DEVICES S.A. DE C.V.
ave. fomento industrial l9 m3
parque industrial del norte
reynosa, tamaulipas, c.p, 88736
MX  88736
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key24701333
Report Number3003442380-2026-99797
Device Sequence Number16062187
Product Code FPA
UDI-Device Identifier05705244020030
UDI-Public05705244020030
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/01/2024
Device Model NumberMMT-943A
Device Lot Number5362155
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2026
Initial Report FDA Received Date03/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-