Unomedical hereby submits this initial report as part of its complaint remediation activities conducted under a capa/fda action plan.This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this initial information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged based on the review completed on 25-feb-2026 and investigation completed on 11-mar-2026 this medical device report (mdr) is being submitted as the initial report.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions h11: investigation summary this investigation has been updated because the malfunction code changed fromtubing connector detached from infusion set (cannula part/base piece) toleak between tubing and site connector - detachment / significant wetness., which requires additional activities not included in the original investigation dated 04-sep-2023.The original investigation remains valid and has not been modified.As part of this reassessment, the following elements were added to align with current investigation requirements complaint investigation results electronic quality management system (eqms) search: a query was run in the eqms on 11/mar/2026 against "lot number" "5388361" and similar malfunction codes leakage from connection between the tubing connector and the infusion set (cannula part/base piece), tubing detached from tubing-tubing connector, tubing connector is cracked, split, deformed, leakage may occur, leak between tubing and site connector - detachment / significant wetness.The review confirmed that lot 5388361 and the identified failure mode are not associated with any non-conformance report (ncr) or (capa) of the same or similar nature.Similar complaints search: a query was run in the eqms on 11/mar/2026 against "lot number" criteria equal "5388361" and similar malfunction code leakage from connection between the tubing connector and the infusion set (cannula part/base piece), tubing detached from tubing-tubing connector, tubing connector is cracked, split, deformed, leakage may occur, leak between tubing and site connector - detachment / significant wetness.The number of complaints is 1.The complaint numbers is complaint (b)(4).Device history record (dhr) review: the lot 5388361 was manufactured according to work instruction (wi) version 45 and manufactured in the m14 on 28/may/2022 with a total of (b)(4) units.The sub-assembly: cannula the lot 5384304 was manufactured according to the wi- version 8 and manufactured in the lc04, on 17-may-2022, with a total of (b)(4) units.The sub-assembly: gluing of tubing the lot 5386000 was manufactured according to the wi- version 54 and manufactured in the spot 05 - 5c06 on 27-may-2022, with a total of (b)(4) units.Conclusion: dhr review supports compliance with manufacturing and quality requirements; no issues noted.Visual evidence review: no photo was provided to support visual confirmation of the reported issue.Conclusion: unable to perform visual verification; assessment based on available documentation only.Capa determination assessment - conclusion summary of complaint investigation: based on the investigation, no further investigation is required.The record will be closed and monitored through tracking and trending per wi- (monthly trips and alerts).Complaint unconfirmed: following investigation the reported failure could not be confirmed for this complaint.
|