| Model Number |
7322 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
An email was received regarding a cadd administration set with item 21-7322-24 and lot 6140818.It was reported that the pump had alarmed ¿completed.¿ the remaining reservoir volume had been 60 ml, and the length of infusion had been 46 hours.The air detector had been off, the upstream sensor had been off, and the lock level had been set to 2.The pump had not been used to prime the extension tubing; the tubing had been primed manually.The event occurred while in use with a patient at the patient's home and there was no patient harm.
|
| |
|