|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.The product was returned to depuy synthes for evaluation.Visual analysis of the returned device revealed that the t-handle was found broken at the threaded rod.The broken fragment was returned for evaluation.No other damage was found.The fracture characteristics are consistent with torsional overload from over tightening the clamp while securing the patella.No evidence of material or manufacturing defects were found that could have contributed to the initiation or propagation of the fracture to failure.Per the attune fixed bearing porous knee surgical technique pages 72-74, the surgeon is instructed to use one hand to position the clamp centralized over the implant and in the correct alignment, while using the other hand to slowly rotate the t-handle clockwise while observing uniform seating of the implant until full seating is achieved.Full seating is achieved when the metal surface is in direct contact with the bone, however, when the patella is fully seated, there may be a slight gap between the polyethylene back and the surface of the bone which is acceptable.Using two hands or excessive torque may indicate an issue with the bone preparation or seating alignment and could cause damage to the instrument (as observed with the returned broken t-handle) or bone.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the att patella clamp por would have contributed to a device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|