| Model Number |
G47308 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the "wall" of an aurous centimeter sizing catheter "cracked" below the hub of the catheter during a transjugular intrahepatic portosystemic shunt (tips) procedure.The user noted a "weird rating", when the pressure monitor was turned on.The user noted bodily fluid leaking, and the crack was identified.The hub was not wiped with any cleaning solution.A power injector was not used; rather manual hand injection was performed.The procedure was completed by using a different unspecified catheter.A section of the device did not remain inside the patient's body.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 catalog #: n5.0-35-100-p-10s-pig-csc-20-01 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|