| Catalog Number |
5C4483 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The devices have been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the twist clamp of two (2) minicap transfer sets remained partially opened and could not be closed.This was identified during use of the devices for peritoneal dialysis therapy.The transfer sets were replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h3, h6 and h11.H11: two (2) actual devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Leak, clear passage, and clamp function testing were performed and no issues were observed.The twist clamps fully engaged in the closed position.Torque engagement test was performed; the engage and disengage force to open and close the twist sleeve was acceptable and met specification.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|