| Model Number |
2120 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during testing the pump had a detached dso seal.There were no patient involvement and no harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was returned for analysis.Visual inspection found a lifted (dso) downstream occlusion seal.Functional and visual tests were performed, and the reported issue was duplicated.The cause of the reported issue was a damaged (dso) downstream occlusion seal.The downstream occlusion seal was replaced.The software was updated as a precaution.The service history review identified no indication that the complaint was related to any service performed on the device during the review period.
|
| |
|