Model Number M00583100
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event or Problem Description
It was reported to boston scientific corporation that a truetome jag 44 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat cholangitis performed on (b)(6) 2026.During the procedure, it was reported that upon the incision the sphincterotome cutting wire suddenly ruptured.The procedure was completed with another of the same device.No further information has been obtained despite good faith efforts.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.
 
Event or Problem Description
It was reported to boston scientific corporation that a truetome jag 44 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat cholangitis performed on (b)(6) 2026.During the procedure, it was reported that upon the incision the sphincterotome cutting wire suddenly ruptured.The procedure was completed with another of the same device.No further information has been obtained despite good faith efforts.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: investigation results the device was not returned for analysis despite good faith efforts; therefore, a technical analysis could not be performed.With all the available information, boston scientific concludes that the reported event of wire break could not be confirmed.The device was not returned for analysis despite good faith efforts.Without analysis of the actual device, there is a lack of objective evidence, or descriptive conditions of the event required to determine a probable root cause of the event.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETOME JAG 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key24706583
Report Number3005099803-2026-01119
Device Sequence Number18780785
Product Code KNS
UDI-Device Identifier08714729858645
UDI-Public(01)08714729858645(17)270814(10)37215530
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K122203
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00583100
Device Catalogue NumberM00583100
Device Lot Number0037215530
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2026
Supplement Date Received by Manufacturer03/30/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date04/23/2026
Date Device Manufactured08/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexMale
-
-