• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOLIFE AB - 3005445717 LUCAS 3 CHEST COMPRESSION SYSTEM; COMPRESSOR, CARDIAC, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOLIFE AB - 3005445717 LUCAS 3 CHEST COMPRESSION SYSTEM; COMPRESSOR, CARDIAC, EXTERNAL Back to Search Results
Model Number LUCAS
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker evaluated the customer's device and was unable to duplicate the reported issue.Stryker replaced the device's compression module.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that their device alarmed and paused unexpectedly while it was running.As a result, the device may be unable to be used for automatic cpr, if needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCAS 3 CHEST COMPRESSION SYSTEM
Common Device Name
COMPRESSOR, CARDIAC, EXTERNAL
Manufacturer (Section D)
JOLIFE AB - 3005445717
scheelevagen 17
ideon science park
lund SE-22 3 70
SW  SE-223 70
Manufacturer (Section G)
JOLIFE AB - 3005445717
scheelevagen 17
ideon science park
lund SE-22 3 70
SW   SE-223 70
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key24707127
Report Number3005445717-2026-00103
Device Sequence Number18751704
Product Code DRM
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K173553
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLUCAS
Device Catalogue Number99576-000063
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2026
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/27/2026
Supplement Date Received by Manufacturer05/28/2026
Initial Report FDA Received Date03/26/2026
Supplement Report FDA Received Date06/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-