Model Number 26055CB
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not returned at the time of mdr submission.The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
The following information was received: during tcrf loop broke inside patient, the bipolar cut was then activated causing a spark and burn/crack to the ceramic beak on sheath.Dark charred marks were noted on fibroids however customer swapped set and carried on resection as normal, at the end of the procedure the dark marks were no longer present.The broken part was retrieved from the patient.No negative impact in state of health reported.The following information was received: during tcrf loop broke inside patient, the bipolar cut was then activated causing a spark and burn/crack to the ceramic beak on sheath.Dark charred marks were noted on fibroids however customer swapped set and carried on resection as normal, at the end of the procedure the dark marks were no longer present.The broken part was retrieved from the patient.No negative impact in state of health reported.Cross reference mdr 9610617-2026-00674.
 
Additional Manufacturer Narrative
The device in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH RESECTOSCOPE SHEATH,22CHARR
Common Device Name
INNER SHEATH RESECTOSCOPE SHEATH,22CHARR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key24708742
Report Number9610617-2026-00673
Device Sequence Number10467262
Product Code HIH
UDI-Device Identifier04048551275597
UDI-Public04048551275597
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K221893
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26055CB
Device Catalogue Number26055CB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2026
Supplement Date Received by Manufacturer04/16/2026
Initial Report FDA Received Date03/30/2026
Supplement Report FDA Received Date04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1