• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752200
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A pharmacist reported that loss of suction occurred during the phacoemulsification phase during cataract surgery with intraocular lens implantation.The surgery was completed on the same day by using the second cassette.There was no patient impact.
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.Product evaluation was not able to be performed since the reported product was not received at the investigation site for evaluation.The investigation conducted a non-conformance review of the reported batch number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.The root cause of the customer's complaint could not be established as a product has not been received, and the condition of the product could not be verified.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.No further investigative or manufacturing actions are warranted at this time due to the product not being returned to the manufacturing site for testing.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24711373
Report Number1644019-2026-01741
Device Sequence Number5740065
Product Code LRO
UDI-Device Identifier00380657522002
UDI-Public(01)00380657522002(17)270331(11)250422(10)17D5W3
Combination Product (Y/N)N
Initial Reporter CountryRE
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752200
Device Lot Number17D5W3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/05/2026
Supplement Date Received by Manufacturer04/28/2026
Initial Report FDA Received Date03/30/2026
Supplement Report FDA Received Date05/19/2026
Date Device Manufactured04/22/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-