| Catalog Number |
1009-00 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for investigation.Therefore, we could not conclusively determine the root cause of the reported incident.Previous investigations into this issue have found the failure was determined to be centering hoop to wire welding error along with operator adjustment error.During the welding process, this centering hoop was likely not properly aligned against its retainer slot.Capa: 2022-025 was previously implemented to address this issue, and no further action is required at this time.The device involved was manufactured after the capa was implemented.Although the issue occurred post-implementation, the overall occurrence of this issue has decreased since the capa was put in place.The production and traceability record for the device was reviewed; no issues were found during manufacturing or packaging that would cause or contribute to the reported event.All quality control tests were completed successfully and met specification.We have not received any other complaints of a similar nature for devices from this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that the lemaitre valvulotome had been used twice, and upon removal from the body, it was observed that the blade would not retract into the sheath.No additional complications or patient injury were reported.It is unknown if the device was checked prior to use.
|
| |
|